Below, you can find quick answers to frequently asked questions on the Eclipse Aided Cortical module.
For more detailed information, please refer to the Interacoustics Academy complete guide or the Aided Cortical quick guide.
The Aided Cortical module comes with three sets of speech-like stimuli:
The ManU-IRU stimuli have been developed by the Interacoustics Research Unit (IRU) and Manchester University. The ManU-IRU stimuli are aligned with the International Speech Test Signal (ISTS) and are excellent in their frequency specificity.
The HD-Sounds are updated versions of the natural speech sounds created by Dr. Harvey Dillion, Ph.D. where the updated stimuli are made more frequency specific.
The Ling sounds: /m/, /u/, /a/, /i/, /sh/, /s/ are shortened versions of the original Ling-6 (HL) sounds. The Ling sounds have been shortened to 100 ms to make them better suited for cortical responses.
The Aided Cortical module can be used as a supplement to the eABR test to evaluate a patient’s audibility using their hearing implant.
However, aided cortical testing should not be performed during CI surgery instead of eABR, and eABR should still be used as a way of measuring CI-electrode function and auditory nerve integrity.
This is a decision which needs to be made on a case-by-case basis depending on the hearing loss and expected audibility for each individual patient, as well as what the clinical question is.
Cortical responses are generated bilaterally and are not limited to one side only because neural pathways cross. For this reason, cortical responses are recorded between the right channel and the vertex electrode, and contralateral responses are recorded from the left channel and the vertex electrode.
No, there are no age limitations. However, it is worth noting that the waveform morphology will look very different on infants when compared to waveforms from adults.
When dealing with infants, having the vertex electrode on the top of the head is difficult because of the skull not being fused. Having to shave the patient’s head to improve impedances is also another challenge.
Furthermore, Visram et al. (2023) mention that placing the vertex electrode on the true vertex versus high forehead yields no significant response differences [1].
Balanced impedances for all electrodes at or below 5 kiloohms are optimal.
This is because the Aided Cortical module uses a detector specifically designed to detect aided cortical responses.
The Fmpi value will show after 10 sweeps.
The Fmpi detector calculates the individual's response to the stimuli, so the response confidence is given in percentage to show the individual patient’s response to the stimuli.
To use the free field speaker, you must connect the jack plug to the ‘Bone Outp.’ socket. The Aided Cortical module uses the Bone Outp. plug in the setup as well.
SpRefL is an abbreviation for Speech Reference Level and is used in the Aided Cortical module because the speech sounds are presented with reference to the ISTS.
By default, three intensities are available for use:
It is still possible to add intensities in the system setup.
You can perform aided cortical testing on both cochlear-implant candidates and cochlear-implant users. However, you may see artifacts on the waveform with cochlear implants.
In these cases, it is important to review the morphology of the waveform alongside the Fmpi value when making an interpretation to assess whether the Fmpi value could be related to the presence of artifacts or a genuine cortical response.
The Fmpi cannot distinguish an artifact generated by a cochlear implant from a cortical response. To allow for easier identification of an artifact, setting the low pass display filter to none may assist in this.
Changing the low pass recording filter in the system setup to 100 Hz may help the recording if you consistently see large artifacts.
The Sound Field Analysis opens the Sound Field Analyzer, which you can use to compensate for any acoustic changes in the test room. For example, if you move the patient further away from the initial calibration, the Sound Field Analysis compensates for this increase in distance.
By using and placing the provided ambient microphone at the place of testing, all stimuli will be measured and will display deviations from the calibrated values. If this is the case, press ‘adjust to target’, which will compensate for the new test conditions. This only takes a few seconds.
No, the Sound Field Analysis is not a calibration. The Sound Field Analysis is a compensation that is referenced to a calibrated target.
A yearly calibration of the Eclipse should still be performed.
In the EP calibration software, it is possible to reduce the intensity of the Sound Field Analyzer.
If this happens, it is because the system is not calibrated. The system must be calibrated before use, and this must be done at the test location.
For response present, the Fmpi detector will aid as a stop criterion, as will the waveforms displayed. You can split waveforms in A and B buffers as known from ABR and display both channels separately. The waveform reproducibility feature known from ABR is also available.
For response absent, the system has a built-in residual noise calculator. Noise targets will be different than those used in ABR because cortical responses contain more noise.
No, the Aided Cortical module uses its own calibration, making it work independently of other modules. So, you can plug in the bone conductor after you have done your aided cortical testing and do testing with a BC protocol.
In aided cortical testing, the patient is awake and is supposed to be engaged in some sort of activity. This makes it difficult to reach the same residual noise values in aided cortical testing as in ABR.
The patient should be engaged in a silent activity. For example, silent toys or silent movies. Toys or movies with sound could mask the stimuli presented from the speaker.
Stimulus color is another way of visualizing your waveforms. By default, the color of each waveform depends on what ear state you have chosen. For example, your waveform will be blue if you have chosen the state Aided Left (AL).
By choosing stimulus color, the color of your waveforms is based on the stimulus and not by the ear. Stimulus colors are set by default.
You can do this in the calibration software under ‘Aided Cortical general setup’.
[1] Visram, A. S., Stone, M. A., Purdy, S. C., Bell, S. L., Brooks, J., Bruce, I. A., Chesnaye, M. A., Dillon, H., Harte, J. M., Hudson, C. L., Laugesen, S., Morgan, R. E., O'Driscoll, M., Roberts, S. A., Roughley, A. J., Simpson, D., & Munro, K. J. (2023). Aided Cortical Auditory Evoked Potentials in Infants With Frequency-Specific Synthetic Speech Stimuli: Sensitivity, Repeatability, and Feasibility. Ear and hearing, 44(5), 1157–1172.