How to perform TEOAE on the Titan without using a PC

By Interacoustics

Prepare the equipment

  1. Turn on Titan by pressing the R or L button.
  2. Choose the test protocol from the list using the up & down arrow keys and then press Select.
  3. To use a different protocol than the one selected, press Protocol and then repeat step 2 above.
  4. Select the ear for testing.

Test environment

The ideal test environment is a quiet room. High ambient background noise adversely affects OAE measurements.

Prepare the baby

The baby should be sleeping or in quiet relaxed state. Sucking, blinking, crying or movement may affect testing.

Place transducer

Place the probe in the ear - the light will turn green when a good seal is obtained.

Run test

Press the START button on the Titan or press the shoulder box button.

Results

PASS result REFER result

PASS - when the criteria for a pass are reached, PASS is displayed in green above the measurement. REFER - when the criteria for a pass are not reached within the measurement time, REFER is displayed in amber above the measurement.

INCOMPLETE - if the test is stopped before a PASS or REFER is generated by the system, INCOMPLETE is displayed above the measurement indicating that the full test was not completed

Presenter

Interacoustics

Available Training

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-28

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-28

How to perform DPOAE on the Titan without using a PC

Learn how to perform handheld DPOAE with the Titan.

Interacoustics2026-10-28

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-28

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-28

How to perform ABRIS on the Titan without using a PC

Learn how to perform handheld automated ABR with the Titan.

Interacoustics2026-10-28

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-28

Selecting the Correct Ear Tip

Learn how to select the correct ear tip for testing with the Titan.

Interacoustics2026-10-28

Pressurized TEOAE

Learn how to perform pressurized transient evoked otoacoustic emission (TEOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Pressurized DPOAE

Learn how to perform pressurized distortion product otoacoustic emission (DPOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-28

How to perform TEOAE on the Titan using a PC

Learn how to perform PC-based TEOAE with the Titan.

Interacoustics2026-10-28

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-28

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-28

OAE Testing with PE Tubes

Learn how to perform OAE testing with PE-tubes using the Titan.

Interacoustics2026-10-28

Daily System Check

Learn how to perform a daily system check of your Titan.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?