OAE Testing with PE Tubes

By Interacoustics

This quick guide describes the procedure when wishing to perform an OAE test on a patient with PE-tubes.

When performing an OAE test the in-ear detection ensures that that the probe is properly inserted in the ear, as well as calculates the correct stimuli level for the testing.
I the patient has PE-tubes inserted, the Titan will in many cases indicate out-of-ear, because of the hole in the eardrum. In this case, follow the procedure as described below.

Prepare the instrument

  1. Turn on Titan by pressing the R or L button.
  2. Select the desired OAE test protocol from the dropdown list.
  3. Select the ear for testing.

Test environment

The ideal test environment is a quiet room. High ambient background noise adversely affects OAE measurements.

Prepare the patient

The patient should be sleeping or in quiet relaxed state. Talking, swallowing, blinking, crying or movement may affect testing.

Place transducer

Place the probe in the ear - the light will turn green when a good seal is obtained.

Run test

When the probe check indicates out-of-ear, hold down the start button for 5 seconds. Now, the screen on the Titan will indicate 'Forced Start'. When the button is released, the test will start and the text '*711 coupler calibration used' is stated at the top of the graph.

Picture 1: After holding down the 'Start' for 5 seconds, 'Forced start' appears in top of the screen.
Picture 2: A measurement performed using forced start, will be indicated by the text '*711 coupler calibration used', also in the saved session.

Presenter

Interacoustics

Available Training

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-28

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-28

How to perform DPOAE on the Titan without using a PC

Learn how to perform handheld DPOAE with the Titan.

Interacoustics2026-10-28

How to perform TEOAE on the Titan without using a PC

Learn how to perform handheld TEOAE with the Titan.

Interacoustics2026-10-28

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-28

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-28

How to perform ABRIS on the Titan without using a PC

Learn how to perform handheld automated ABR with the Titan.

Interacoustics2026-10-28

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-28

Selecting the Correct Ear Tip

Learn how to select the correct ear tip for testing with the Titan.

Interacoustics2026-10-28

Pressurized TEOAE

Learn how to perform pressurized transient evoked otoacoustic emission (TEOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Pressurized DPOAE

Learn how to perform pressurized distortion product otoacoustic emission (DPOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-28

How to perform TEOAE on the Titan using a PC

Learn how to perform PC-based TEOAE with the Titan.

Interacoustics2026-10-28

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-28

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-28

Daily System Check

Learn how to perform a daily system check of your Titan.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?