How to perform ABRIS on the Titan without using a PC

By Interacoustics

Prepare the equipment

  1. Turn on Titan by pressing the R or L button.
  2. Choose the test protocol from the list using the up & down arrow keys and then press Select.
  3. To use a different protocol than the one selected, press Protocol and then repeat step 2 above.
  4. Select the transducer (only if more than one connected to the preamplifier).
  5. Select the ear for testing.

Test environment

The ideal test environment is a quiet room where lights and other electronic equipment are turned off.

Prepare the infant

1. Patient state

The infant should be sleeping or in quiet relaxed state. Sucking, blinking, crying or movement may affect testing.

2. Skin preparation

Typically, no skin preparation is required.
If the infant's skin is oily, the electrode placement sites should be cleaned using alcohol wipes and/or skin preparation gel.
Use a conductive gel with the electrodes to improve impedances.

Place electrodes

Place surface electrodes using the desired montage (mastoid or nape).

Connect cables

Connect the cables from the PreAmplifier to the respective surface electrodes.

Check impedance

Impedances are indicated by the green/amber dots on the infant as well as in numerical format on screen.

When the dot is amber, this means the impedance is poor (> 40kΩ). In this instance, it may be necessary to clean the skin and replace the surface electrode.

NB. Testing is possible when impedances are poor. This may, however, affect test time and measurements.

Place transducer

Place the probe or insert earphones in the infant's ear/s or place the headset over the infant's ears.

The probe light will turn green when a good seal is obtained.

Run test

Press the START button on the Titan or press the preamplifier button.

Patient noise (EEG)

After starting the test, the patient noise/EEG bar is displayed at the top of the screen.

If the patient noise goes above the black line, try to calm the infant to reduce movement, crying, sucking etc.

If the infant appears calm and the patient noise is not ideal, stop the test and select a protocol with a higher rejection level (if available).

If the patient noise/EEG signal is still not ideal, refer to the EEG troubleshooting section below.

page2image52386560 page2image52394816

PASS - when the result reaches 100% for the pass criteria, PASS is displayed in green above the measurement.
REFER - when the result does not reach 100% for the pass criteria within the measurement time, REFER is displayed in amber above the measurement.
INCOMPLETE - if the test is stopped before a PASS or REFER is generated by the system, INCOMPLETE is displayed above the measurement indicating that the full test was not completed.

EEG troubleshooting

When the patient noise/EEG signal is poor, and the patient is calm, try the following:

  • Turn off (if possible) all electronic devices/sources, such as lights, computers, mobile phones etc.
  • Ensure that the Titan battery is not being charged (via the cradle) during testing
  • If impedances were initially poor, clean the skin/use conductive gel and replace the electrodes

Presenter

Interacoustics

Available Training

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-28

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-28

How to perform DPOAE on the Titan without using a PC

Learn how to perform handheld DPOAE with the Titan.

Interacoustics2026-10-28

How to perform TEOAE on the Titan without using a PC

Learn how to perform handheld TEOAE with the Titan.

Interacoustics2026-10-28

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-28

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-28

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-28

Selecting the Correct Ear Tip

Learn how to select the correct ear tip for testing with the Titan.

Interacoustics2026-10-28

Pressurized TEOAE

Learn how to perform pressurized transient evoked otoacoustic emission (TEOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Pressurized DPOAE

Learn how to perform pressurized distortion product otoacoustic emission (DPOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-28

How to perform TEOAE on the Titan using a PC

Learn how to perform PC-based TEOAE with the Titan.

Interacoustics2026-10-28

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-28

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-28

OAE Testing with PE Tubes

Learn how to perform OAE testing with PE-tubes using the Titan.

Interacoustics2026-10-28

Daily System Check

Learn how to perform a daily system check of your Titan.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?