How to Verify Directional Hearing Aids

By Interacoustics


 

 

Directionality Test using Real Ear Measurements

  1. From your patient management system, Noah or OtoAccess®, launch the Affinity or Suite.
  2. Launch the REM440 module by clicking on the REM tab.
  3. Prepare the patient for Real Ear Measures.
  4. Calibrate the probe tube.
  5. If fitting an open hearing instrument, perform Calibrate for open fit.
  6. Ensure that the correct Ear is selected.
  7. To activate the Directional Test, go to: Menu | Setup | Show/hide protocols | US | Standard Directionality. Insert the checkmark inside the box in front of Standard Directionality. This will ensure that the Directionality test is available from the list of protocols drop down menu (f choosing US country defaults during the Affinity or Callisto software installation, the Directionality test will be available by default)
  8. Choose the preferred Stimulus in the corresponding dropdown list.
  9. Change the input level using the slider if desired.
  10. Click START and follow the instructions on screen
  11. Place the client in front of the loudspeaker as shown on the picture. Click OK and front microphone curve will be recorded. New instructions will immediately appear. Turn the client around and click OK.

 

How to add the directionality test to an existing protocol

  1. Go to Menu | Setup | REM440 Setup.
  2. Choose a protocol to copy (ex: Adult Aided Response) from the drop down list.
  3. Click New to create a new protocol.
  4. Click Yes when the software asks if you want to copy the currently selected protocol (If selecting NO, you will have to create the protocol from scratch, which is beyond the scope of this document. For more info, please refer
    to the Instructions for Use document, Additional Info document or Quick Guide on Protocol Setup )
  5. Rename the Protocol.
  6. Add the Directionality test to the existing protocol.
  7. Edit the settings if desired by clicking on the Settings button
  8. Adjust the Settings if desired .

Presenter

Interacoustics

Available Training

How to perform frequency lowering verification

In this Affinity 2.0 quick guide, you'll learn how to determine if you should apply frequency lowering and how to verify the frequency lowering feature.

Interacoustics2026-10-28

BKB-SIN speech-in-noise test

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with the Interacoustics Affinity 2.0 audiometer. Read more.

Interacoustics2026-10-28

Binaural Real Ear Measurements

Learn how to perform binaural real ear measurements (REM) with the Affinity 2.0.

Interacoustics2026-10-28

How to perform FM transparency verification

In this Affinity 2.0 quick guide, you'll learn how to perform FM transparency verification. Read more.

Interacoustics2026-10-28

Verifying Ear Level FM Systems

Learn how to verify ear level FM systems with the Affinity.

Interacoustics2026-10-28

AC440 Protocol Setup

There are many different demands and wishes regarding how an audiometer should both function and appear depending on the specific situation and national standards. One of the greatest advantages of the AC440 module is the flexibility, enabling you to tailor the system according to your specific preferences in the so-called “test protocols”. This article describes basic functionality on how to work with the AC440 Protocols.

Interacoustics2026-10-28

Print Reports

Learn how to create customized print templates for the Affinity 2.0, which can be linked to individual protocols for quick printing.

Interacoustics2026-10-28

Hearing Instrument Testing

Learn how to perform hearing instrument testing with the Affinity 2.0 device.

Interacoustics2026-10-28

Real-Ear-to-Coupler Difference

Learn how to measure the real-ear-to-coupler difference (RECD) with the Affinity 2.0 device.

Interacoustics2026-10-28

QuickSIN Test

Learn how to perform the quick speech in noise (QuickSIN) test with the Affinity 2.0 device.

Interacoustics2026-10-28

Basic Audiometry

Learn how to do basic audiometry with the Affinity 2.0 device.

Interacoustics2026-10-28

TEN Test in AC440

Learn how to perform the threshold equalizing noise (TEN) test with the Affinity 2.0 device.

Interacoustics2026-10-28

How to Use the SKS10 Skull Simulator

Click here to learn what a skull simulator is and how you can use skull simulators to measure the performance of bone anchored hearing devices.

Interacoustics2026-10-28

Adult Hearing Aid Verification

Learn how to verifiy hearing aids in adults with the Affinity 2.0 device, using real ear measurements (REM).

Interacoustics2026-10-28

Visible Speech Mapping

Learn how to perform visible speech mapping with the Affinity 2.0 device.

Interacoustics2026-10-28

Speech in Quiet (SIQ) Test

Learn how to perform the speech in quiet (SIQ) test with the Affinity 2.0 device.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?