High Frequency Audiometry

By Interacoustics

What is high frequency audiometry?

High frequency audiometry (above 8 kHz) is performed using the same procedure as normal air conduction audiometry. High frequency audiometry is helpful when testing hearing impairments caused by ototoxicity, noise exposure, and acoustic traumas or in the assessment of patients with tinnitus. The frequency area is more susceptible to the effects of external factors such as medication and loud noises relative to the low and mid frequencies.

Pure tone audiogram measured for the right ear. dB HL is a function of frequency. Hearing thresholds have been recorded from 0.25 kHz to 12.5 kHz.

Required equipment

  • A response button.
  • High frequency audiometry headset (e.g. DD450) and a high frequency license installed.

Note high frequency audiometry is only available if the optional high frequency license is installed on the instrument. If the license is not installed or if the headset is not calibrated for high frequencies the HF phone button does not appear on the audiometer and the HF button and HFz button will be dimmed in Diagnostic Suite.

Test procedure

  1. To perform HF Audiometry, select the Tone and HF Phone; this will activate the HF Audiometry function .
  2. Choose between HF and HFz; shows the full range of audiometric test frequencies, whereas HFz only shows the high frequency range or in Diagnostic Suite.
  3. Use the high frequency headset to perform a high frequency The audiometry is performed in the same way as a normal air conduction audiometry would be performed.

Presenter

Interacoustics

Available Training

How to install and use Diagnostic Suite with your AS608

In this video, you'll learn how to install and use Diagnostic Suite with your AS608. Watch the video here.

Interacoustics2026-10-28

Diagnostic Suite: How to create a print template

In this video, you'll learn how to create a print template with Diagnostic Suite. Watch the video here.

Interacoustics2026-10-28

Diagnostic Suite: How to share a print layout design

In this video, you'll learn how to share a print layout design with Diagnostic Suite. Watch the video here.

Interacoustics2026-10-28

Short Increment Sensitivity Index

Learn how to perform the short increment sensitivity index (SISI) test in the Diagnostic Suite software.

Interacoustics2026-10-28

Tone Decay Test

Learn how to perform the tone decay test in the Diagnostic Suite software.

Interacoustics2026-10-28

Pediatric Noise

Learn about the pediatric noise stimulus, which is available with in the Diagnostic Suite software.

Interacoustics2026-10-28

Speech Audiometry

Learn how to perform speech audiometry in the Diagnostic Suite software, including results interpretation.

Interacoustics2026-10-28

Pure Tone Audiometry

Learn how to perform pure tone audiometry in the Diagnostic Suite software.

Interacoustics2026-10-28

Weber Test

Learn how to perform the Weber test in the Diagnostic Suite software, including results interpretation.

Interacoustics2026-10-28

Quick Speech in Noise (QuickSIN)

Learn how to perform the quick speech in noise (QuickSIN) test in the Diagnostic Suite software, including results interpretation.

Interacoustics2026-10-28

Threshold Equalizing Noise (TEN) Test

Learn how to perform the threshold equalizing noise (TEN) test in the Diagnostic Suite software, including results interpretation.

Interacoustics2026-10-28

Masking Level Difference (MLD)

Learn how to measure the masking level difference (MLD) in the Diagnostic Suite software, including results interpretation.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?