Weber Test

By Interacoustics

What is the Weber test?

The Weber test distinguishes between conductive and sensor neural hearing loss through use of a bone conductor. Use the indications to show where the tone is perceived.

Required equipment

  • AD629 audiometer
  • A bone conductor

Weber test procedure

An image of Test procedure

  1. Press Tests - Weber to enter the Weber test screen.
  2. Place the bone conductor on the forehead of the patients and instruct them to tell you if the tones presented are heard better to the Right, Left, Centre or Not heard at all.
  3. Present a tone at a level of 10 dB above the worst BC. You can select whether you want a tone or Warble stimulus pressing the hardkey Tone/Warble on the device.
  4. Await response from the patient and click on the corresponding soft button.

An image of response

Weber test results

If the patient hears the tone better in the poorer ear, the hearing loss is conductive. If the tone is heard better in the better ear, the hearing loss is sensor neural at the given frequency.

Presenter

Interacoustics

Available Training

BKB-SIN speech-in-noise test

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with the Interacoustics AD629 diagnostic audiometer. Read more.

Interacoustics2026-10-28

How to print without a database

Click here to learn about the three options you have for printing without a database with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Short Increment Sensitivity Index

Learn how to perform the short increment sensitivity index (SISI) test with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Audiometric Masking

Learn how to perform audiometric masking with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

High Frequency Audiometry

Learn how to perform high frequency audiometry with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Békésy Audiometry

Learn how to perform the Békésy audiometry test with the AD629 diagnostic audiometer, including the Békésy audiometry types for results interpretation.

Interacoustics2026-10-28

Speech Audiometry

Learn how to perform speech audiometry with the AD629 diagnostic audiometer, including results interpretation.

Interacoustics2026-10-28

Pure Tone Audiometry

Learn how to perform pure tone audiometry with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Stenger Test

Learn how to perform the Stenger test with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Alternate Binaural Loudness Balance Test

Learn how to perform the alternate binaural loudness balance (ABLB) test with the AD629 diagnostic audiometer, including results interpretation.

Interacoustics2026-10-28

Hughson-Westlake Procedure

Learn how to perform the Hughson-Westlake procedure with the AD629 diagnostic audiometer.

Interacoustics2026-10-28

Threshold Equalizing Noise (TEN) Test

Learn how to perform the threshold equalizing noise (TEN) test with the AD629 diagnostic audiometer, including results interpretation.

Interacoustics2026-10-28

Quick Speech in Noise (QuickSIN) with AD629

Learn how to perform the quick speech in noise (QuickSIN) test with the AD629 diagnostic audiometer, including results interpretation.

Interacoustics2022-02-11

Looking for clinical training?

Are you looking for more advanced and clinical training?