TEOAE Probe Test

By Interacoustics

Probe performance is crucial to TEOAE test results. We recommend that you conduct a probe test at the beginning of each day before starting to test on patients to ensure that the probe is functioning correctly.

 

  • Before conducting the probe test, ensure that the probe tip is clean and free of wax and/or debris.
  • Always conduct the probe test in a quiet test environment.
  • Only use the recommended cavity for testing. Using a different type of cavity may either not detect probe faults or may incorrectly indicate a faulty probe.

 

Performing the Probe Test

  1. Select the Probe Test TEOAE protocol on the handheld device or in the PC software.


  2. Insert the probe tip without an ear tip attached into the Probe Test Cavity* or the 0.5cc** cavity provided with the Titan. Press the Start button and let the test run until it stops (approx. 30 seconds). Do not stop the test manually.
  3. If the probe is functioning correctly, none of the TE bands will have a checkmark above them when the test ends. It is possible to continue with daily testing.


  4. If error messages appear during testing or if one or more of the TE bands has a checkmark above it at the end of the test, the probe test has failed. Check and clean the probe tip for wax or debris and redo the probe test. If the probe test fails a second time, the Titan must not be used to test on patients. Contact your local distributor for assistance.

Note: Failure of the daily probe test also indicates that TEOAE measurements performed since the last successful probe test may be invalid and patients may need to be retested. Hence, the importance of performing the probe test daily.

For more information about the probe test, please refer to the Titan Instructions for Use Manual.

* A specifically designed Probe Test Cavity will be available for use shortly. Until such time, please use the 0.5cc cavity provided with Titan. **The 0.5 cc cavity simulates the impedance of neonate and adult ears to an acceptable level for the daily Probe Test. We discourage the use of smaller cavities for the daily check in neonatal screening as such a cavity is not representative of a neonate ear due to soft tissue in the ear canal.

Presenter

Interacoustics

Available Training

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-28

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-28

How to perform DPOAE on the Titan without using a PC

Learn how to perform handheld DPOAE with the Titan.

Interacoustics2026-10-28

How to perform TEOAE on the Titan without using a PC

Learn how to perform handheld TEOAE with the Titan.

Interacoustics2026-10-28

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-28

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-28

How to perform ABRIS on the Titan without using a PC

Learn how to perform handheld automated ABR with the Titan.

Interacoustics2026-10-28

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-28

Selecting the Correct Ear Tip

Learn how to select the correct ear tip for testing with the Titan.

Interacoustics2026-10-28

Pressurized TEOAE

Learn how to perform pressurized transient evoked otoacoustic emission (TEOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Pressurized DPOAE

Learn how to perform pressurized distortion product otoacoustic emission (DPOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-28

How to perform TEOAE on the Titan using a PC

Learn how to perform PC-based TEOAE with the Titan.

Interacoustics2026-10-28

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-28

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-28

OAE Testing with PE Tubes

Learn how to perform OAE testing with PE-tubes using the Titan.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?