How to Identify a Perforated Eardrum

By Interacoustics

About the author

The information presented here is based on clinical examples as well as modeled patterns. Text and accompanying absorbance sketch is authored by Navid Shahnaz, PhD, Aud, Associate Professor of Audiology in the School of Audiology & Speech Sciences at the University of British Columbia (UBC).

 

Caveats

Firstly, it must be noted that a perforated ear is easiest recognized by the flat curve and high equivalent ear canal volume of the traditional tympanogram. Similarly, the 3D absorbance measure allows recognizing that the measures are constant over the pressure range.

Secondly it should be noted, that recordings on ears with perforations will vary between patients - the shown Absorbance patterns for smaller and larger perforations shown here are sketched examples only. Please see Susan Voss' article referenced below for more details.

 

Absorbance characteristics to look for

The Absorbance generally increases compared to normal ears, and mostly so below 2 kHz as the middle ear cavity absorbs most of the energy. The increase in Absorbance is most pronounced with the smallest(!) perforations. The bump of higher absorbance around 1 kHz shown in this sketch can be very pronounced with a very small perforation and may shift to higher frequencies as the size perforation increases. With larger perforations, the low frequency Absorbance gets closer and closer to normal absorbance levels, but might remain above the Absorbance level the ear would have exhibited without the perforation.

 

Sketched example of an absorbance pattern
Sketched example

 

a 3D example of a normal reading
3D example of normal

a 3D example of perforation
3D example of perforation

 

Consequences of probe fit

All Absorbance measures need to have a good probe fit to be reliable. Evaluating Absorbance as provided by the 3D Tympanometry test ensures that a reasonably air tight probe seal was accomplished, as the air pressure sweep would not have been performed otherwise. In addition to an air tight probe seal, a deeper rather than a shallower probe insertion ensures the most accurate Absorbance measures. Shallow insertions tend to provide more elevated Absorbance readings at lower frequencies. This is somewhat similar to normal Tympanometry measures that are also influenced by probe fit and probe insertion depth.

 

Suggested reading

Effects of Middle-Ear Disorders on Power Reflectance Measured in Cadaveric Ear Canals, Voss, Susan E., Merchant, Gabrielle R.,Horton, Nicholas J., Ear & Hearing. 33(2):195-208, March/April 2012.

Acoustic Immittance Measures, Basic and Advanced Practice, 2013, Lisa Hunter, Phd, FAAA, Navid Shahnaz, PhD, Aud. (C), Plural Publishing. ISBN10: 1-59756-437-0, ISBN13: 978-1-59756-437-3.

Presenter

Interacoustics

Available Training

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-28

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-28

How to perform DPOAE on the Titan without using a PC

Learn how to perform handheld DPOAE with the Titan.

Interacoustics2026-10-28

How to perform TEOAE on the Titan without using a PC

Learn how to perform handheld TEOAE with the Titan.

Interacoustics2026-10-28

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-28

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-28

How to perform ABRIS on the Titan without using a PC

Learn how to perform handheld automated ABR with the Titan.

Interacoustics2026-10-28

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-28

Selecting the Correct Ear Tip

Learn how to select the correct ear tip for testing with the Titan.

Interacoustics2026-10-28

Pressurized TEOAE

Learn how to perform pressurized transient evoked otoacoustic emission (TEOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Pressurized DPOAE

Learn how to perform pressurized distortion product otoacoustic emission (DPOAE) testing with the handheld Titan tympanometer.

Interacoustics2026-10-28

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-28

How to perform TEOAE on the Titan using a PC

Learn how to perform PC-based TEOAE with the Titan.

Interacoustics2026-10-28

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-28

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-28

OAE Testing with PE Tubes

Learn how to perform OAE testing with PE-tubes using the Titan.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?