Free Field Adjustment

By Interacoustics

The purpose of this document is to provide a quick guide for the Free Field Adjustment feature found in the AC440 Audiometry module in the Affinity Suite. This quick guide will focus on how to use this function.

Introduction

Free field measurements are really valuable when it comes to understanding your patients hearing function via speech tests or even applying aided testing. However they can be difficult to apply correctly in constantly changing environments, whether this is because you travel between different sites or there are changes to your clinic room. We realise this and have implemented the Free field Adjustment feature in the AUD module.

This feature allows you to refresh your FF speaker calibration when moving your equipment between different environments or simply to reconfirm your speaker calibration in a fixed setup. This brings confidence when testing as it ensures your stimulations are accurate to base your diagnosis from.

Preparation

  1. Ensure that you have the FF Calibration Adjustment microphone and FF speakers plugged into your Affinity Compact. The FF Cal socket has been highlighted in the image below.
    Affinity Compact sockets, with red square around socket labeled ambient cal mic.
  2. Launch your Affinity Suite and go into the AC440 Audiometry Module. Click the FF Adjustment icon which can be found in the left panel.
    Icon with speaker and the words 'Free field adjustment'.
  3. You will now see the below screen
    Message box that states the following: The microphone should be pointing upwards at head height in the position that the patient would sit for the test. In the lower right corner, there is a 'Start' button.
    You are required to hold the FF Cal. Mic upright in the position that the patient should sit during testing to complete this process
  4. Press the 'Start' button in the Free field adjustment window

    When the FF adjustment process is in progress you will see the bar begin to fill and there will be several stimuli played from the speaker(s) to the Ambient-cal mic.

    The process is fully automated and when completed the system will apply corrections based on the room acoustics to ensure that the correct stimulus can be achieved when performing free field tone or speech audiometry.

    This process removes the need to have your instrument fully re-calibrated for new environments.

    Note: Please keep the environment silent during this process and any interfering noise will affect the calibration levels set or cause the process to fail.

    Note: This process does not overwrite your existing calibration files, it simply adds a correction to them so that the correct stimulus is applied at the reference point in your new environment. On closing the suite this correction is removed.

    Note: You will need to re-perform this process for every patient tested.

  5. Once this process has successfully completed the Free field adjustment window will state 'completed', see example image below


  6. Proceed to perform FF Audiometry or Speech testing as you desire.

Presenter

Interacoustics

Available Training

AC440 Protocol Setup

Learn how to set up protocols in the Affinity Compact's audiometry module.

Interacoustics2026-10-28

How to perform FM transparency verification

In this Affinity Compact quick guide, you'll learn how to perform FM transparency verification. Read more.

Interacoustics2026-10-28

How to perform frequency lowering verification

In this Affinity Compact quick guide, you'll learn how to determine if you should apply frequency lowering and how to verify the frequency lowering feature.

Interacoustics2026-10-28

BKB-SIN speech-in-noise test

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with the Interacoustics Affinity Compact audiometer. Read more.

Interacoustics2026-10-28

Unboxing Affinity Compact

In this unboxing video, you'll see the various hardware included when purchasing the Affinity Compact from Interacoustics. Click here to watch the video.

Alan Thaastrup Mikkelsen2026-10-28

Performing an RECD Measurement

Learn how to perform coupler fittings with the Affinity Compact.

Dennis Mistry2026-10-28

How to Perform Real-Ear Measurements

Learn how to perform real-ear measurements with the Affinity Compact.

Dennis Mistry2026-10-28

Affinity Compact: Features and functions

This video provides an overview of the Affinity Compact's features and functions.

Dennis Mistry2026-10-28

Affinity Compact: A product introduction

This video provides an introduction to and overview of the Affinity Compact.

Interacoustics2026-10-28

How to Use Affinity Compact's Probe Placement Indicator

Learn how to use Affinity Compact's Probe Placement Indicator in this short, step-by-step guide. Read the guide here.

Rasna Kaur Mistry, BSc (Hons) Audiology2026-10-28

Customized Printout Setup

Learn how to customize printouts with the Affinity Compact.

Interacoustics2026-10-28

Real-Ear-to-Coupler Difference

Learn how to measure the real-ear-to-coupler difference (RECD) with the Affinity Compact.

Interacoustics2026-10-28

QuickSIN Test

Learn how to perform the quick speech in noise (QuickSIN) test with the Affinity Compact.

Interacoustics2026-10-28

TEN Test in AC440

Learn how to perform the threshold equalizing noise (TEN) test with the Affinity Compact.

Interacoustics2026-10-28

REM440 Protocol Setup

Learn how to set up protocols in the Affinity Compact's real-ear measurement module.

Interacoustics2026-10-28

Visible Speech Mapping

Learn how to perform visible speech mapping with the Affinity Compact.

Interacoustics2026-10-28

Looking for clinical training?

Are you looking for more advanced and clinical training?