Hearing aid verification using fitting systems has become very popular since the introduction of digital hearing aids.
Real-ear measurements (REM) are the main way in which a clinician can measure the hearing aid's effects objectively and independently to gain a better idea of the hearing aid's performance and suitability (Aazh and Moore, 2007; Mueller, 2001).
There are two methods of performing REM: Aided Response and Insertion Gain.
Let's begin by explaining the difference between them.
Aided Response is a method of verifying the hearing aid's output in dB SPL. It involves measuring the total sound achieved at the probe microphone. This includes the signal, the patient's own ear canal effect and the hearing aid's amplification (Pumford and Sinclair, 2001; Mueller, 2001).
Because Aided Response is working with the total sound, no REUR/REUG needs to be measured as it is being included in all measurements that are being taken.
It is also for this reason that the gain target curves will not change after measurement of the REUG/REUR because regardless of the patient's personal differences, the total effect should be the same outcome.
You can still choose to include REUR/REUG to ensure correct tube insertion (Dillon, 2012).
Insertion Gain is a method of verifying the hearing aid's output in dB gain. Whilst still using the measured sound at the probe microphone, the dB gain view only shows the amplification applied by the hearing aid (Pumford and Sinclair, 2001).
Therefore, the effects of the signal and the patient's own ear canal amplification are removed to focus purely on what the hearing aid needs to achieve the target output signal (Mueller, 2001).
Real Ear Insertion Gain equals Real Ear Aided Response minus Real Ear Unaided Response.
As Insertion Gain is a calculation, it is a requirement that the REUR/REUG exists ahead of making any measurements with the hearing aid enabled.
People have been encouraged and educated to perform REM using a certain method.
Both methods still yield the same outcome but they may be showing different things.
Historically, both methods have been suggested for different fitting algorithms (BSA, 2007), with Insertion Gain being most suitable for NAL and Aided Response being most suitable for DSL.
However, both algorithms now have compatibility to be performed in each way (BSA, 2018).
Insertion Gain only shows the hearing aid's amplification, meaning it is difficult to understand the full audibility provided when working in this way.
Aided Response shows the complete effect achieved in the ear from all components, which helps to understand the overall effect but can make it difficult to understand solely what one of the components (i.e. hearing aid alone) is contributing.
It also gives greater visibility regarding the total effect of the sound achieved within the patient's dynamic range and the proximity to the patient's threshold and uncomfortable levels (Bentler et al, 2016).
Remember that the probe-tube microphone placement for the Aided Response method requires a deeper insertion.
This is because your objective here is to measure the sound level close to the eardrum, normally within 5 mm (Dillon, 2012; Pumford and Sinclair, 2001; BS ISO 12124, 2001).
Overall, you need to understand what it is you want to see.
Most clinicians are focused on the hearing aid, but the overall performance in the patient's ear is the main effect of their efforts.
Aided Response is seen to be easier to understand from an audibility perspective, therefore it may simplify your efforts when explaining the hearing aid's effect to your patient.
Aided Response is also a much easier method to use when starting to work with REM (BSA, 2018).
Insertion Gain is a very objective way to look at verification of hearing aids as it singles out just one of the components contributing to the solution achieved in the patient's ear canal.
Should you be interested in the hearing aid's performance alone, then this is the best method to use.
1. Aazh H. and Moore B.C. (2007). The Value of Routine Real Ear Measurement of the Gain of Digital Hearing Aids. J Am Acad Audiol, 18, 653–664.
2. Bentler, R., Mueller, H.G., Ricketts, T.A. (2016). Modern Hearing Aids Verification, Outcome Measures, and Follow-Up. ISBN13: 978-1-59756-482-3.
3. British Society of Audiology. (2007). Guidance on the Use of Real Ear Measurement to Verify the Fitting of Digital Signal Processing Hearing Aids.
4. British Society of Audiology. (2018). Guidance on the verification of hearing devices using probe microphone measurements.
5. BS ISO 12124. (2001). Acoustics: Procedures for the Measurement of real-ear acoustical characteristics of hearing aids. Geneva, International Standards Organisation.
6. Mueller, H. G. (2001). Probe Microphone measurements: 20 years of progress. Trends in amplification, 5(2), 35.
7. Pumford, J. and Sinclair, S. (2001). Real-Ear Measurement: Basic Terminology and Procedures. AudiologyOnline.
Subscribe to our newsletter and receive news on new products, seminars and much more.
By signing up, I accept to receive newsletter e-mails from Interacoustics. I can withdraw my consent at any time by using the ‘unsubscribe’-function included in each e-mail.
Click here and read our privacy notice, if you want to know more about how we treat and protect your personal data.