How to prepare the patient for oVEMP testing

10 January 2024
10 mins
Video

Transcript

Before you start testing, instruct the patient to turn off all their electronic devices in order to minimize noise.

Now, look in your patient's ear to ensure no excess cerumen is present.

Clean the patient's skin on the forehead, between the eyes, and under each eye.

Place your electrodes, mount your cables.

Remember, the left cable should be connected to the electrode under the right eye.

And the right cable should be connected to the electrode under the left eye.

When the cables are placed correctly, test the impedances.

Impedances below five kiloohms are acceptable.

It is important that there is no more than two kiloohms of imbalance in impedances between electrodes.

Insert the transducers and instruct the patient to keep a constant upward gaze during testing without moving their head.

You do not need the EMG monitor for oVEMP.


Presenter

Interacoustics

Popular Academy Training

Interacoustics - hearing and balance diagnosis and rehabilitation
Copyright © Interacoustics A/S. All rights reserved.