Versions | DPOAE test options | TEOAE test options |
---|---|---|
Screener+ | 2 Fixed protocols – DP 2s & DP 4s 2 - 5 kHz 65/55 dB SPL 6 dB SNR 3 out of 4 frequencies for a pass Test time of 2 or 4 sec per frequency |
2 Fixed Protocols – TE 32s & TE 64s 1.5 - 4 kHz 80 dB pe SPL 4 dB SNR 3 out of 6 frequencies for a pass Test time of 32 or 64 sec (max test time) |
Screener DPOAE | 2 Fixed protocols – DP 2s & DP 4s 2 - 5 kHz 65/55 dB SPL 6 dB SNR 3 out of 4 frequencies for a pass Test time of 2 or 4 sec per frequency |
n/a |
Screener TEOAE | n/a | 2 Fixed Protocols – TE 32s & TE 64s 1.5 - 4 kHz 80 dB pe SPL 4 dB SNR 3 out of 6 frequencies for a pass Test time of 32 or 64 sec (max test time) |
Standard DPOAE | 1 screening + 4 customizable protocols 1.5 - 12 kHz 12 frequency test protocols available Customizable stimulus intensity, SNR, test time, number of frequencies required for a pass |
n/a |
Standard TEOAE | n/a | 1 screening + 2 customizable protocols User selectable frequencies between 0.7 - 4 kHz 80 dB pe SPL Customizable SNR, test time, number of frequencies required for a pass |
Clinical | 1 screening + 4 customizable protocols 1.5 - 12 kHz 12 frequency test protocols available Customizable stimulus intensity, SNR, test time, number of frequencies required for a pass |
1 screening + 2 customizable protocols User selectable frequencies between 0.7 - 4 kHz 80 dB pe SPL Customizable SNR, test time, number of frequencies required for a pass |